44. Replace the open/short combination with the 50 ohm termination (supplied
in the calibration kit).
46. Connect the "open" end of the open/short combination to the analyzer test
port 2.
48. Connect the "short" end of the open/short combination to the analyzer test
port 2.
49. press ~~~~~~~~~.
50. Connect the 50 ohm termination to the analyzer test port 2.
52. When the analyzer displays PRESS 'DONE' IF FINISHED WITH STD(s) , press
Wait for the message COMPUTING CAL COEFFICIENTS to disappear from the
analyzer display before proceeding to the next step.
53. Connect the equipment as shown in F'igure 2-30.
Figure 2-30. Transmission Calibration 'I&t Setup
H P 0753E
N E T W O R K A N A L Y Z E R
CABLE
24 INCH
System Verification and
Performance Tests
2-63