- Average: setting of the number (1, 4, 16, 32 or 64) of readings to
calculate the arithmetic average (AVG).
When setting Average = 1, the readings are not averaged and the
current measurements are displayed.
In the example, the average value of the last 16 field readings will be
displayed. The "Avg" counter shows the progressing of the average
calculation, updating according to the "Reading rate" time interval setting.
The message AVG: OK will be displayed shortly after completion of the
averaging process, then followed by the indication of the preset averaging
Avg: 16 .
Then the process continues in moving Average mode: of the 16 readings
block, the older is discarded and the latest is added at the speed of the
Reading rate setting.
The process is restarted from the beginning when setting a different
Average value.
When toggling from TOT to XYZ mode the Average is reset to re-
calculate the field values for each single axis.
When in HOLD mode, the new Average value is displayed in the main
window and is updated only when the HOLD is released.