L3Harris narda PMM EP-600 User Manual page 23

Table of Contents

Advertisement

1.10 Specifications EP-603 This condition applies to all specifications:
TABLE 1-4 Specifications of the electric field probe PMM EP-603
Frequency range
Level range
Overload
Dynamic range
Linearity
Resolution
Sensitivity
Flatness
Isotropicity
Sensors
X/Y/Z reading
Battery reading
Temperature reading
Internal data memory
Battery
Operation time
Recharge time
Dimensions
Weight
Operating temperature
Software for PC
Optical fiber connector
Tripod adapter
• The operating ambient temperature range must be -10° to 50 °C.
300 kHz – 18 GHz
0.17 – 170 V/m
> 350 V/m
60 dB
0.4 dB @ 50 MHz/0.3 – 170 V/m
0.01 V/m
0.17 V/m
3 – 8200 MHz 1.4 dB
1 – 12000 MHz 2.4 dB
0.6 – 18000 MHz 3.8 dB
(With frequency correction OFF)
0.3 – 18000 MHz 0.4 dB
(Typical with frequency correction ON)
0.4 dB (0.2 dB typical @ 50 MHz)
Six monopoles
Simultaneous sampling of the components
10 mV res.
0.1 °C res.
Serial number
Date calibration
Calibration Factor
SW release.
Panasonic ML621S 3V 5mA/h
100 h @ 0.4 S/sec 28 Hz filter
60 h @ 3 S/sec 28 Hz filter
48h for maximum autonomy
21 mm sphere
16 mm sensor
53 mm overall
23g including FO weight (1m)
-10° - +50°
YES
HFBR-0500
¼ - 20 UNC female
rechargeable Li-Mn
General Information
1-9

Advertisement

Table of Contents
loading

Table of Contents