5.5.10 Settings menu
5.5.10.1 Settings
Commands:
- Settings: set Average and Filter
- Preference: select program skin and enable the function Minimized UI
- ? (Info): software and product information
Settings options:
- Filter: the EP-600/601/602/603/604 internal A/D Sigma-Delta converters
feature digital filters to improve resolution and sensitivity, as well to
reducing interferences.
The filters can be selected according to the application requirements: from
the faster (F1) to the slower (F8).
To set the required filter click on the corresponding line, which turns blue.
The selected filter will be displayed in the main window.
When selecting a new filter while in HOLD mode, the filter setting
indication in the main window will be updated only after toggling to
RUN mode.
Filter(Hz): indication of the filter characteristics
Rejection to mains: some filters feature notch filtering at mains frequency
to reducing interference
Settling time (ms): time required for getting a complete reading
Max Sample rate (S/s): A/D conversion speed
Min Reading Rate (s): min. time interval between readings of electric field
WinEP600 and SetAddEP600 operating instructions
5-21