The EP-600/601/602/603/604 field probe is allocated in a spherical plastic
3.7 PMM EP-600/601/
housing with 6 orthogonal monopoles that allow for immediate identification
602/603/604 operation
of the electric field vectors (axis).
The PMM EP-600/601/602/603/604 is supplied from an internal battery that
can be recharged by the charge EP600 CHARGER supplied with.
One full charge cycle before operation is recommended to obtain the
maximum autonomy.
The PMM EP-600/601/602/603/604 can be switched ON by shortly
pressing the pushbutton, after that the LED is sequentially turned on with
green, red and blue indications as a test for the same; then the LED will
blink red, this meaning the EP-600/601/602/603/604 is ready for the
operation.
The PMM EP-600/601/602/603/604 cannot be switched OFF by the
pushbutton.
The PMM EP-600/601/602/603/604 automatically turns OFF when:
- after 180 sec. the fiber optic has been disconnected or the communication
with PC is not established; use the setting command #00e n* (see chapter
6) to set the time before the EP-600/601/602/603/604 auto-switches off.
The command #00e n* is available with Firmware release 1.12 or
higher
- When use the setting command #00! (see chapter 6).
- when the battery voltage is below 2.05V (the battery voltage is displayed
by the software WinEP600).
- when the Software WinEP600 is closed
Do not expose the probe to a field higher than the max. allowed.
Field strengths exceeding the allowed may cause severe probe
damage whether it is connected or not, either turner ON or OFF.
The fiber optic must be connected or disconnected by holding it by
the connectors only. Pulling the fiber optic cable may cause severe
damages to the PMM EP-600/601/602/603/604 and to the fiber optic
cable itself.
Dust and dirt must be prevented to be in between the optical
connections.