L3Harris narda PMM EP-600 User Manual page 54

Table of Contents

Advertisement

Fix
Fix
Fix
Blinking
Fast Blinking
Slow Blinking
Fast Blinking
* The led status, of the blue conic charger, is based on current consumption.
When the current is very low, the charger interprets this as "no load" on it and the LED is fixed
green.
In case the battery is fully charged the led on the charger is immediately green without blinking
because the current absorbed by the probe, is very low.
** The red status of the led happens when there is an overloading or a short circuit on the pin of
the blue conic charger.
In this case you have to remove the ac/dc charger from the blue cone to give it the possibility to
restore itself, just some seconds are more than enough
*** Available from June 2018, for new products or after repair & calibration service
4-6
EP600 CHARGER
Table 4-2 EP600 CHARGER Led status - Start up phase
The EP600 CHARGER is supplied but without load (battery not
Green
present or disconnected or completely full) *.
Orange
Charge ended after 10 minutes of no load (led fix green) ***
The charging circuit is overloaded (the output is OFF for 30
Red
seconds before attempting a new charging cycle).
The central contact has been shorted **
Table 4-3 EP600 CHARGER Led status - Charger phase
Green
The battery is under charge
Red
Charge ended for maximum time limit reached [ 60 hours ]
Charge ended for maximum time limit reached without 12 hours
Orange
of stabilisation.
Charge ended for maximum time limit reached with 12 hours of
Orange
stabilisation.
The complete charging cycle is of 48 hours; the charge is completed
when the Led of EP600 CHARGER blinks orange.

Advertisement

Table of Contents
loading

Table of Contents