1.4 Specifications EP-600 This condition applies to all specifications:
TABLE 1-1 Specifications of the electric field probe PMM EP-600
Frequency range
Level range
Overload
Dynamic range
Linearity
Resolution
Sensitivity
Flatness
Isotropicity
Sensors
X/Y/Z reading
Battery reading
Temperature reading
Internal data memory
Battery
Operation time
Recharge time
Dimensions
Weight
Operating temperature
Software for PC
Optical fiber connector
Tripod adapter
• The operating ambient temperature range must be -10° to 50 °C.
100 kHz – 9.25 GHz
0.14 – 140 V/m
> 300 V/m
60 dB
0.4 dB @ 50 MHz/0.3 – 100 V/m
0.01 V/m
0.14 V/m
1 – 150 MHz 0.8 dB
0.5 – 6000 MHz 1.6 dB
0.3 – 7500 MHz 3.2 dB
(With frequency correction OFF)
0.3 – 7500 MHz 0.4 dB
(Typical with frequency correction ON)
0.5 dB (0.3 dB typical @ 50 MHz)
Six monopoles
Simultaneous sampling of the components
10 mV res.
0.1 °C res.
Serial number
Date calibration
Calibration Factor
SW release.
Panasonic ML621S 3V 5mA/h
100 h @ 0.4 S/sec 28 Hz filter
60 h @ 3 S/sec 28 Hz filter
48h for maximum autonomy
21 mm sphere
16 mm sensor
53 mm overall
23g including FO weight (1m)
-10° - +50°
YES
HFBR-0500
¼ - 20 UNC female
rechargeable Li-Mn
General Information