WinEP600 and SetAddEP600 operating instructions
Normally a fast field measurement is required for faster response and
better stability when the field probe controls as feedback for a preset field
value the radiated power of a system composed by antenna and RF
generator.
However, as faster the measurements as higher the noise is, with
consequent reduction of sensitivity.
Slower measurements increase sensitivity but increase the response time
and the power consumption.
The PMM EP-600/601/602/603/604 features eight different filters to help
finding the most suitable combination of the a.m. parameters for each
specific application:
Power consumption
Very low
Very low
Low
Average
Average
Average
High
Very high
In normal operation setting the filters F4 - F5 may offer a good
compromise in terms of power consumption, sensitivity, settling time
and rejection @ 50Hz.
Sensitivity
Rejection
@ 50Hz
Low
-----
Average
Low
Good
------
High
Good
High
High
High
Very high
Very high
good
Very high
High
Rejection
@ 60Hz
low
------
-------
Very high
High
Good
Good
High