5.5.5 XYZ / TOT
Button for selecting the x - y – z (XYZ) or Total mode (TOT).
Max. display resolution: 0.001 V/m.
Click on the XYZ button to toggle to Total mode.
If the field value measured is outside the nominal level range, the following
messages are displayed:
Ovr : the field level is higher than 110% of the nominal maximum level (550
V/m for EP601).
! : the field level is between 100% and 110% of the nominal maximum level
(from 500 to 550 V/m for EP601). The symbol is close to the value.
* : the field level is lower than nominal minimum level. The symbol is close
to the value.
0.5 V/m per EP601 on the total;
0.5 / √¯ 3 per EP601 on the single axis.
Low : the field level is lower than 1/15 of the nominal minimum level.
0.5 / 15 per EP601 on the total
0.5 / (15 x √¯ 3) per EP601 on the single axis.
The field value outside the nominal level range could be unreliable.
WinEP600 and SetAddEP600 operating instructions
5-13