Troubleshooting
Error Messages
When an error condition occurs, the SkanStacker 300 stops operation and an
alarm signal is heard. The LED Ready light flashes and a message is shown in
the display.
The sensor check is performed once when the
tiate a wash. During the wash, the SkanStacker 300 does not check the sen-
sors.
Error
E-01 Bad Vacuum
E-02 Air Pressure
Pressure High
Low
Verify Setup
E-03 MicroPlate
No Plate In Lift
Insert Plate
37
Cause
Low vacuum at the
vacuum outlet when
starting wash.(Vac-
uum must be less than
900 mbar).
The
key is
[Start]
pressed for wash or
rinse/prime. The actual
air pressure is not
equal to the set air
pressure value.
The
key has
[Start]
been pressed for wash
or rinse/prime but no
microplate is in the
plate holder.
key is pressed to ini-
[Start]
Things to Try
Verify vacuum pump is
switched on.
Verify tubing connected
properly.
Verify caps on reservoir
bottles and waste collec-
tion vessel are tight.
Verify vacuum tubing
free from restrictions or
obstructions.
Verify pump protection
filter is not closed or
filled with fluid.
Low: Probably a leak.
Check that all reservoir
caps are tightened.
High: Due to changed air
pressure setting. The previ-
ously used wash program
had a higher air pressure
setting. Open reservoir cap
to release excess air pres-
sure.
Insert a microplate.
SkanStacker 300 User Guide