An all-inclusive automatic microplate washer
Table of Contents

Advertisement

Wash Step Selections
Each program can have up to 16 wash steps. There are seven wash step
selections available for each step. Press
use
arrow keys to change settings. Press
Wash Step
Rinse Time
Disp Volume
Soak/Pause
Wash volume
Wash time
Aspirate
End Wash
Table 1: Wash Step Selections
SkanStacker 300 User Guide
Description
Set time in seconds from 0.1 to 10 seconds. Liquid will
be dispensed from "Rinse" port. Fluid is agitated during
rinse.
Set dispense volume from 100-350 µl. Liquid will be
dispensed from "Inlet" port.
Set soak/pause time from 5 seconds to 2 minutes 50
seconds.
Set volume of wash from 100-900 µl. Liquid will be dis-
pensed from "Inlet" port.
Set wash time from 0.1 to 10 seconds. Liquid will be dis-
pensed from "Inlet" port. Fluid is agitated during wash.
Set aspiration time from 2.0 - 10 seconds. Lift velocity is
varied with selected aspirate time. The lift velocity is
automatically set so that at the last second of the aspira-
tion, the aspirate probes are at the bottom of the well.
Aspirate times longer than 5 seconds will have the same
lift velocity as the 5-second setting.
Aspirate time = 2.0 seconds, lift velocity approximately
1 second.
Aspirate time = 5 seconds, lift velocity approximately 4
seconds.
Set for last step of wash.
!! IMPORTANT:
Exit/Save must be selected after
program setting completed.
Software Configuration
to enter selection, and
[Enter]
to select.
[Enter]
To save changes,