Chapter 4 Troubleshooting; Error Messages - Molecular Devices StakMax User Manual

Microplate handling system for the spectramax paradigm multi-mode detection platform
Hide thumbs Also See for StakMax:
Table of Contents

Advertisement

Troubleshooting

Error Messages

The following error messages can appear in the Log View frame in the
StakMax Software window.
Table 4-1 Troubleshooting table
Error
Code
200
201
202
203
204
205
206
207
5012878 A
Error Message
Error: A problem was
encountered during
Reset.
Error: Call Molecular
Devices Technical
Support at +1-800-
635-5577.
Error: A problem
occurred during a
motion.
Error: A problem
occurred with the X
axis.
Error: A problem was
encountered during
Reset.
Error: A problem
occurred during a
motion.
Error: A Problem
occurred with the
EPROM.
Error: A problem
occurred during a
motion.
Description
Homing routine was
not able to find a
state change on the
home flag,
sensor/flag problem
likely source.
Timer resource was
not available,
apparently not used
in code.
Stall occurred during
motion, encoder and
step counts exceeded
maximum threshold.
X Axis motion did not
cause state change
on edge detect switch
—poorly gripped
plate, plate not
present.
Can't move axis if it
hasn't been homed.
Can't move axis if it's
already in motion.
Generic EPROM error. Call Molecular Devices
Carriage drop off
exceeds maximum
travel of X Axis.
4
Solution
Press the
Power/Reset button.
Call Molecular Devices
Technical Support
+1-800-635-5577.
Check for object
blocking movement
and then re-align.
Check to make sure
plate is properly
gripped and then re-
align. If error
persists, call
Molecular Devices
Technical Support
+1-800-635-5577.
Press the
Power/Reset button.
Press the
Power/Reset button.
Technical Support
+1-800-635-5577.
Re-align the
instrument.
105