SkanStacker 300 User Guide
Description
Use to exit program. User able to save or reject data
before exiting.
Program name selected up to 5 characters.
Select each character by using the
keys.
Press
[Enter]
position.
Characters can be capital or lower-case letters,
digits, or special characters such as brackets,
asterisks, colons, etc.
Delete program by moving
blank space.
Pressure set for internal air pressure pump.
Set to any value from 0.0 to 0.6 bar
(approximately 9 psi).
Pressure is changed to adjust the inlet wash
liquid flow.
To disable the pressure pump, select 0.0 bar.
For normal wash, select 0.25 bar (approximately
4 psi).
Adjustment can be made to fine tune the dispensed vol-
ume using dispense or wash functions. The instrument is
set at a default of 1.00 which can handle water and up to
0.02% Tween. If a viscous material is used, increase the
dispense volume by increasing the factor (range 0.01 to
2.50).
Software Configuration
to move to the next character
arrow key to
arrow
16