CAUTION
Select this menu item to alter the values of the polynomial coefficients contained in the formula used
to calculate moisture concentration. Press
X = 0, 1, 2, 3, D, E. Enter the new coefficient value when the "New Value?" prompt is displayed.
Enter the new limit value when the "New Value?" prompt is displayed. Press
and display the next coefficient; press

Labels

This menu item is used to change the text and format of the first three display lines. Highlight
the label to be changed, press
changed then use the up and down arrows to scroll through the list of available ASCII characters.
Press
to select a character. Note that the character ~ indicates a placeholder for a numerical
enter
display including decimal points and negative signs; a total of eight places is allowed. Do not erase
the leading ~ placeholder; add or remove a ~ placeholder from the right-hand side only. When
finished, press
enter
The following labels may be altered:
Alter Moisture Label
(Line_X_Flag = 1: Held Moisture During Equilibration, Cal, and Zero)
Alter Pressure Label
(Line _X_Flag = 10: Cell Pressure)
Alter Temperature Label Default: ~~~~ Degree Cent
(Line_X_Flag = 11: Cell Temperature in °C)
Alter Flow Label
(Line_X_Flag = 6: Flow Rate) Note that internal flowmeter is necessary for this data to be
available.
Alter Current Label
(Line_X_Flag = 9: Live Moisture including Equilibration, Cal, and Zero)
5-8 2850 Moisture Analyzer
The polynomial coefficients used in the moisture concentration calculations vary from
instrument to instrument and are determined during factory calibration. With the
exception of Coefficient D, these values should never be changed unless factory service
is performed or a major component is changed.
cancel
cancel
. Use the left and right arrows to highlight the character to be
enter
to save and
to quit.
cancel
Default: ~~~ ppm Vol H2O
Default: ~~~~~ psi Cell Prs
Default: ~~~~~ sccm Flow
Default: ~~~~~~ Current H2O
to roll through the variables PolyCoeff_X where
as required to quit.
to save change
enter