Test/Config Main Menu

• View Configuration

Selecting this menu item allows software configuration values to be viewed only. Current software configu-
ration should be compared periodically to the factory software configuration data sheet to verify that critical
system values have not been changed. Note that it is normal to see a slight change in values for Moist_Span
and PolyCoeff_D as these may be changed during calibration and zero cycles.
To view configuration values, press
Config Main Menu. The following parameters and current settings will be displayed in the order shown.
Moist_Span
Moist_Gen
Press_Span
Press_Offset
Normal_Press
Temp_Span
Temp_Offset
• Set Time of Day
At "Enter New Time" prompt, enter four digit time of day using the 24 hour clock convention. Note that it
is not necessary to enter punctuation between hours and minutes.
• Set Date
At "Enter New Date" prompt, enter six digit date in the form MMDDYY. Note that it is not necessary to enter
punctuation between month, day, and year.
• Set Auto Cal Time
At "Enter New Time" prompt, enter the time of day for the auto-calibration cycle to be run (this becomes the
value of the variable Cal_Time). Enter four digit time using the 24 hour clock convention. Note that it is not
necessary to enter punctuation between hours and minutes.
• Set Auto Zero Time
Not Used. Please contact AMETEK Service at 800-537-6044 for more information on implementing fully
automatic zero cycle.
• System Tests Menu
The Model 2850 Analyzer has numerous self-test routines which can be run periodically as part of regular
maintenance or used to assist in troubleshooting the analyzer. The system tests are discussed in detail in
(Chapter 6): Troubleshooting and Maintenance section of this manual.
• Configure Menu
Both Level 2 and Level 3 Configure Menus are accessible through this menu item. Refer to (Chapter 5):
Software Configuration for more information on the Configure Menus and associated functions.
to display the next parameter; press
enter
Flow_Span
Flow_Offset
Temp_Set_Pt
Propor_Band
Temp_Reset
Convr_Factor
Flags_1
Flags_2
Output2_Flag
Line_1_Flag
PolyCoeff_0
Line_2_Flag
PolyCoeff_1
Line_3_Flag
PolyCoeff_2
Alarm_1_Flag
PolyCoeff_3
Alarm_2_Flag
PolyCoeff_D
Output1_Flg
PolyCoeff_E
to return to the Test/
cancel
Controller Function 4-9