SMARTLINE® RM60 (FMCW)
Measurement accuracy
Resolution
Repeatability
Accuracy
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Temperature
Pressure
Relative air humidity
Target
Operating conditions
Temperature
Temperature
Temperature
Temperature
Ambient temperature
Storage temperature
Flange temperature
Pressure
Pressure
Pressure
Pressure
Operating pressure
Other conditions
Other conditions
Other conditions
Other conditions
Dielectric constant (εr)
Vibration resistance
Protection category
Maximum rate of change
34-VF-25-31 iss.4 GLO December 15 US
1 mm / 0.04¨
±1 mm / ±0.04¨
±3 mm / ±0.12¨, when distance < 10 m / 33 ft;
±0.03% of measured distance, when distance > 10 m / 33 ft
+20°C ±5°C / +70°F ±10°F
1013 mbara ±20 mbar / 14.69 psia ±0.29 psi
60% ±15%
Metal plate in an anechoic chamber
-40...+80°C / -40...+175°F (according to the temperature limits of the gasket
material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
-40...+85°C / -40...+185°F
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Standard: -50...+150°C / -58...+300°F
Option: -50...+200°C / -58...+390°F
(the process connection temperature must agree with the temperature limits of the
gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
-50...+150°C / -58...+300°F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
-40...+100°C / -40...+210°F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
-1...16 barg / -14.5...232 psig;
subject to process connection used and flange temperature
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
-1...40 barg / -14.5...580 psig;
subject to process connection used and flange temperature
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Standard: -1...40 barg / -14.5...580 psig;
subject to process connection used and flange temperature
≥1.5
IEC 60068-2-6 and EN 50178 (10...57 Hz: 0.075 mm / 57...150 Hz:1g)
IP66/67 equivalent to NEMA type 4X (housing) and type 6P (antenna)
10 m/min / 33 ft/min