Calibration Of Ultramat 23; Ultramat 23 Autocal Calibration Of Zero Point - Siemens Set CEM CERT Operating Instructions Manual

Continuous gas analysis
Hide thumbs Also See for Set CEM CERT:
Table of Contents

Advertisement

Functions
9.1 Calibrating
9.1.2

Calibration of ULTRAMAT 23

The Set CEM CERT system is configured so that the ULTRAMAT 23 analyzer (option
C10, C20, C30) automatically performs an AUTOCAL with ambient air every 24 hours.
This adjusts the zero point and sensitivity of the IR channels. The sensitivity of the O
channel (if present) is also adjusted. The timing for this can be set. (Menu → Calibration
U23 → Press AUTOCAL display)
In addition, the full-scale value must be readjusted every 6 months with calibration gas.
BEFORE calibration of the full-scale value, we recommend carrying out an AUTOCAL
with ambient air. Allow approximately 30 minutes for a complete calibration and take this
into account if you have to guarantee regulatory availability of the AMS or provide half-
hour averages.
9.1.2.1

ULTRAMAT 23 AUTOCAL calibration of zero point

Press the following buttons:
A complete AUTOCAL will now be performed. While this is taking place, the status traffic
light at the top right is orange. As soon as this disappears again, the AUTOCAL is
finished.
50
"Menu"
"Calibration U23"
Operating Instructions, 03/2020, A5E34491309-11
"Start AutoCal"
Set CEM CERT
2

Advertisement

Table of Contents
loading

Table of Contents