Physical setting data
Gas cooler
Monitoring of heated sample gas line -EH1
•
K2 - Monitoring
•
K1 - Control
Temperature setting on probe
Cabinet temperature monitoring -KT2
Cabinet air conditioning
(manufacturer's setting on air conditioning
unit)
Thermostat for frost protection heater -EH10
1)
The setpoint temperature of the heated sample gas line and the probe may differ depending on the acid dew point
of the gas. The maximum setpoint temperature depends on the type of probe and the type of sample gas line. Ob-
serve the information provided by the respective manufacturer.
Time intervals
Calibration
•
ULTRAMAT 23
•
ULTRAMAT 6
Maintenance
Set CEM CERT
Operating Instructions, 03/2020, A5E34491309-11
Setpoint +4 °C, set at factory
Setting on -KT1
T
+140 °C
min
1)
T
+180 ℃
set
1)
T
+180 ℃
set
Tmax +35 °C
15 °C ... 35 °C
8 °C
24 h must be set for the automatic zero point calibration (Thermo
AUTOCAL function activated).
A weekly interval must be provided for the automatic zero and refer-
ence point calibration.
See current certificate
Technical specifications
65