10.5

Emptying the condensate tank

The condensate level in the condensate tank must be regularly checked. If the level is
too high, a maintenance request is displayed via the sensor. The condensate tank must
be emptied when this maintenance request appears, at the latest.
While the condensate tank is being emptied, the measuring system operates without
condensate disposal. The condensate tank must be emptied and reconnected as quickly
as possible, because condensate may otherwise accumulate in the cooling paths of the
cooler and impair measurements.
After reconnection of the condensate tank and the capacitive sensor, the condensate
disposal is fully functional again.
The distance between the capacitive sensor and the condensate tank determines the
sensitivity of the display.
To empty the condensate tank, proceed as follows:
1. Loosen the hose couplings on condensate tank (couplings close automatically when
loosened).
2. Detach the capacitive sensor together with the mounting bracket (level monitoring).
3. Carefully remove the tank.
4. Dispose of the condensate from the tank properly via the yellow cap.
5. Reconnect the tank using the coupling and attach the capacitive sensor together with
the mounting bracket between the tank and tank handle.
10.6

Spare parts/accessories

For spare parts, please refer to the SIOS Process Analytics
(https://support.industry.siemens.com/cs/products?dtp=Manual&mfn=ps&pnid=13613
&lc=en-WW).
Set CEM CERT
Operating Instructions, 03/2020, A5E34491309-11
Service and maintenance
10.5 Emptying the condensate tank
59