Connection
6.1 Mechanical installation
Use the provided screwed glands to connect the zero and calibration gases. Unused gas
inlets must be provided with blanking plugs.
NOTICE
Inability of system to function
Prevent pressure variations and any back pressure at all in the line because this may
cause the analyzers to be damaged. For calibration gas cylinders, a pressure reducer
with a maximum output pressure of 0.5 bar is required.
6.1.2

Condensate outlet

The separated liquid from the sample gas cooler is discharged using hose pumps.
Accumulated condensate produced is collected in a tank intended for that purpose
(optional) or routed directly from the system through PTFE tubes. If the condensate tank
is full, a warning is output on the display. This disappears after the condensate tank has
been emptied. The owner must ensure proper collection and disposal of the condensate
according to environmental regulations.
6.1.3

Gas outlet

6.1.3.1
Gas outlet
Ensure that the exhaust gas line is correctly dimensioned for the number of analyzers.
High pressure variations or any back pressure at all is not permitted at the sample gas
outlet.
30
Operating Instructions, 03/2020, A5E34491309-11