Settings after commissioning

Make the following settings on the analyzers:
• Configure and check the measuring ranges in accordance with EN 15267
• Set the analog outputs to 4 ... 20 mA (retain measured value or update measured
value in the case of AUTOCAL)
• Check the relay contacts:
– R2= Readiness for measurement
– R4 = Function check
– R6 = Maintenance
– R8 = Zero gas
Note
Because these settings are made on the analyzers themselves, always observe the
information contained in the respective operating instructions/manuals.
7.2.3.1

ULTRAMAT 23

Input of calibration gas concentration
Before a calibration can be performed, you must set the calibration gas concentration.
This setting is made directly on the analyzer.
Proceed as follows for this purpose:
1. Press <ENTER> key → Operating menu appears
2. → Select calibration
3. → Select Calibrate IR MR
4. → Select component
5. → Select Set calibration gas setpoints
6. → Input sample gas concentration for MR1 and MR2
7. → To save, press the <MEAS> key
Repeat this procedure for other components as needed.
Set CEM CERT
Operating Instructions, 03/2020, A5E34491309-11
Commissioning
7.2 Commissioning