Functions

9.1

Calibrating

9.1.1

General

It is possible to calibrate the installed analyzers at the analyzer itself. For details on
operating the analyzers during calibration, please refer to the manuals of the analyzers.
These are included with the end customer documentation when the analyzer is delivered
or can be ordered or downloaded via the Siemens Industry Online Support (SIOS) SIOS
Process Analytics
(https://support.industry.siemens.com/cs/products?dtp=Manual&mfn=ps&pnid=13613
&lc=en-WW).
Calibration and zero gases must be connected according to the gas flow diagram before
commissioning. Also make sure that adequate gas is available in the gas cylinders so
that a constant gas flow is ensured.
Note
Incorrect calibration results
After the cabinet is opened, different temperatures between the interior of the cabinet
and the environment may cause a drifting of the measuring instruments, in particular if
the hinged frame with the devices is also swung out. In this case, wait until the
measured values of the analyzer are no longer drifting before starting the calibration.
This can take up to 2 hours, depending on the temperature difference.
Before starting the calibration process, set a constant flow rate of the gases between 72
l/h and 120 l/h on the flow meter of the analyzer. Ensure that this flow value remains
constant throughout the calibration process.
To keep the environmental conditions constant during calibration, we recommend using
the HMI display installed in the cabinet door for the calibration. After calibration is
complete, the analyzer is ready for operation.
As soon as you make a setting or open the cabinet, switch on Maintenance in the menu.
Switch it off again after the calibration.