Technical specifications
System design
Color
Weight
Sheet-steel cabinet/frame
Dimensions (with base)
GRP cabinet
Dimensions (with base)
Explosion protection classification
Degree of protection
Operating position
Sample gas cooler
Calibration
•
ULTRAMAT 23
•
ULTRAMAT 6, OXYMAT 6
Analyzers
1)
A space of 500 mm (20") must be provided on the right or left side for the cable inlet and the connection of the
heated sample gas line.
Climatic conditions
Permissible ambient temperature
•
During operation
•
With heating in sheet-steel cabinet
•
With heating in GRP cabinet
•
With air conditioning
Permissible ambient humidity
The conditions of EN 15267-3 must be observed.
Set CEM CERT
Operating Instructions, 03/2020, A5E34491309-11
RAL 7035
Approx. 160 kg (353 lbs.)
Indoor installation
1)
2100 x 800 x 800 mm; 83" x 31 1/2" x 31 1/2" (H x W x D)
Outdoor installation
1)
2060 x 900 x 800 mm; 81" x 35 1/2" x 31 1/2" (H x W x D)
Installation outside the Ex zone
IP 54 (avoid direct sunlight when installed outdoors)
Vertical
2-stage
Semi-automatic; AUTOCAL freely adjustable up to max. 24-hour in-
terval
Fully automatic
ULTRAMAT 23, ULTRAMAT 6, OXYMAT 6 and SIPROCESS UV600
For further technical details of the analyzers, refer to the correspond-
ing operating instructions. This information is included in section 4 of
the overall documentation with your delivery.
+5 °C ... +40 °C, standard; conforms to EN 15267-3
-5 °C ... +45 °C
-15 °C ... +40 °C
Max. +52 °C
75% RH (relative humidity) annual average, non-condensing
12
63