Maintenance and service

8.1

Safety checks

The following safety checks must be performed every 24 months by a
qualified service engineer:
● Visual inspection of the unit and its accessories for mechanical
damage that might impair operation
● General function check
● Check of the visual and audible indicators
● NC and SFC earth leakage current acc. to IEC 60601
● NC and SFC housing leakage current acc. to IEC 60601
● NC and SFC patient leakage current acc. to IEC 60601
● Laser power measurement with a calibrated measuring instrument in
the range between 0.2 W and 3 W
8 Maintenance and service
8.1 Safety checks
67