Table 10-3. Electromagnetic Immunity for Non-Life-Supporting Systems
Guidance and manufacturer's declaration – electromagnetic immunity Table of IEC 60601-1-2
The INVU by Nuvo™ system is intended for use in the electromagnetic environment specified below. The
customer or the user of the INVU by Nuvo™ Interface should assure that it is used in such an environment.
IMMUNITY test Test level
Immunity to electrostatic discharge (ESD)
IEC 61000-4-2
± 2, 4, 8 kV contact;
± 2, 4, 8, 15 kV air discharge
Conducted Immunity to electrical fast transients/bursts
IEC 61000-4-4
Repetition
frequency:
Test voltage:
Line
description:
Conducted immunity to voltage surges
IEC 61000-4-5
Applied
voltage:
Surge
application
Conducted immunity to voltage dips, short interruptions under minimum 100 VAC, 50 Hz power option
IEC 61000-4-11 Voltage
reduction,
% of Vnom:
Phase
angle:
Duration:
Conducted immunity to voltage dips, short interruptions under minimum 240 VAC 50 Hz power option
IEC 61000-4-11 Voltage
reduction,
% of Vnom:
NV-UMLS-0001
Compliance level
± 8 kV contact discharge;
± 15 kV air discharge;
100 kHz
Repetition
frequency:
2 kV
Test voltage:
Line 1 &
Line
Line 2
description:
±0.5, ±1.0,
Applied
±2.0 kV
voltage:
L1-to-GND;
Surge
L2-to-GND
application:
1. 100%
Voltage
reduction,
2. 100%
% of
3. 30%
Vnom:
4. 100%
1. 0°; 45°;
Phase
90°; 135°;
angle:
180°; 225°;
270°; 315°
2. 0°
3. 0°
4. 0°
1. 10ms
Duration:
2. 20ms
3. 500ms
4. 5000ms
1. 100%
Voltage
reduction,
2. 30%
3. 100%
Invu by Nuvo™User Manual
Electromagnetic environment –
guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
2 kV
Mains AC power should meet
the standards of a typical
domicile, commercial or
100 kHz
hospital environment. AC
Line 1 &
power is applied to INVU by
Line 2
Nuvo™ during the charging
mode only.
±0.5, ±1.0,
AC power should meet the
±2.0 kV
standards of a typical domicile,
commercial or hospital
L1-to-GND;
environment. AC power is
L2-to-GND
applied to INVU by Nuvo™
during the charging mode only.
1. 100%
AC power should meet the
standards of a typical domicile,
2. 100%
commercial or hospital
3. 30%
environment. AC power is
4. 100%
applied to INVU by Nuvo™
1. 0°; 45°;
during the charging mode only.
90°; 135°;
180°; 225°;
270°; 315°
2. 0°
3. 0°
4. 0°
1. 10ms
2. 20ms
3. 500ms
4. 5000ms
1. 100%
AC power should meet the
standards of a typical domicile,
2. 30%
commercial or hospital
3.100%
Appendices
73