10. The number (A) that is displayed in the main dialogue box
at the end of the calibration procedure is the ratio of the
current calibration value to the previous calibration value.
If there has been no change to the load cell, A should be
between 0.99 and 1.01.
11. If the A value is greater than 1.05 or less than 0.95, the
load cells were significantly out of calibration. If this is the
case, the calibration procedure above should be repeated
to ensure that the preload was correct and an accurate
calibration achieved.
12. The load cell calibration procedure must be carried out for
both the X and Y axis.

Snap Image Feature

Located under >Tools>Snap Image
This feature allows the user to capture a single
image during the setup phase. This can be
useful for documenting specimen loading
activity
and
neutral
position.
(non-contact)
beam