Calibrate:voltage[:Level]; Calibrate:voltage:meter:point; Calibrate:voltage:meter[:Level] - ITech IT8800 Series Programming Manual

Programmable dc electronic load
Hide thumbs Also See for IT8800 Series:
Table of Contents

Advertisement

CALibrate:VOLTage[:LEVel]

This command is only effective in calibration mode. According to the reading of
external measuring meter to input the calibration voltage. Please select a
calibration grade before you input a calibration value(using the command
CALibrate:VOLTage:POIN).These constant do not exist in the nonvolatile
memory unless you have sent the command CALibrate:SAVE.
Command Syntax
Parameters
Unit
Examples
Related Commands

CALibrate:VOLTage:METEr:POINt

This command is only effective in calibration mode, is used to set the calibration
point in CV mode.P1, P2 are used in low voltage range, P3, P4 are used in high
voltage range. Only be used to calibrate voltage source and voltmeter.
Command Syntax
Parameters
Examples
Related Commands

CALibrate:VOLTage:METEr[:LEVel]

This command only be effective in calibration mode. According to the reading of
external measuring meter to input the calibration voltage. Before you input the
value, you should firstly select a calibration grade(with command
CALibrate:VOLTage:POINt).All these constant do not exist in the nonvolatile
memory unless you have send the command CALibrate:SAVE to save them.
Command Syntax CALibrate:VOLTage:METEr[:LEVel] <NRf>
Parameters
Unit
Examples
Related Commands
CALibrate:VOLTage[:LEVel] <NRf>
<external reading>
V (volts)
CAL:VOLT 3.2223
CAL:STAT
CALibrate:VOLTage:METEr:POINt <point>
P1 | P2 | P3 | P4
CAL:VOLT:METEr:POIN P2
CAL:STAT
<external reading>
V (volts)
CAL:CURR 3.2223
CAL:STAT
Copyright © ITECH Electronic Co., Ltd.
Calibration Commands
CAL:SAV
CAL:SAV
CAL:SAV
63