Trace:feed:control; Trace:data; Trace:filter; Trace:delay - ITech IT8800 Series Programming Manual

Programmable dc electronic load
Hide thumbs Also See for IT8800 Series:
Table of Contents

Advertisement

Returned Parameters
Related Commands

TRACe:FEED:CONTrol

This command is used to select the buffer control. With NEVer selected,
storage into the buffer is disabled. When NEXT is selected, the storage process
starts, fills the buffer and then stops. The buffer size is specified by the :POINts
command.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands

TRACe:DATA?

When this command is sent and the electronic load is addressed to talk, all the
readings stored in the buffer are sent to the computer.
Query Syntax
Returned Parameters

TRACe:FILTer

This command is used to select whether the data in cache is the data filtered.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters

TRACe:DELay

This command is used to select the delay time for the trigger in cache.
Command Syntax
Parameters
UNIT
*RST Value
Examples
Query Syntax
Returned Parameters

TRACe:TIMer

This command is used to select the interval of the cache.
Command Syntax
Parameters
UNIT
*RST Value
<CRD>
TRAC:POIN
TRACe:FEED:CONTrol <CRD>
NEVer | NEXT
NEVer
TRAC:FEED:CONT NEXT
TRACe:FEED:CONT?
<CRD>
TRAC:FEED
TRACe:DATA?
{<NR3>}
TRACe:FILTer[:STATe] <BOOL>
0 | 1 | ON | OFF
OFF
TRAC:FILT 1
TRACe:FILTer[:STATe]?
<NR1>
TRACe:DELay <NRf>
0 to 3600s | MINimum | MAXimum | DEFault
S (second)
0
TRAC:DEL 1
TRACe:DELay? [MINimum | MAXimum | DEFault]
<NR3>
TRACe:TIMer <NRf>
0.00002 to 3600s | MINimum | MAXimum | DEFault
S (second)
1
Copyright © ITECH Electronic Co., Ltd.
Trace Subsystem
40

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents