Calibrate:current[:Level]; Calibrate:current:meter:point; Calibrate:current:meter[:Level]; Calibrate:voltage:point - ITech IT8800 Series Programming Manual

Programmable dc electronic load
Hide thumbs Also See for IT8800 Series:
Table of Contents

Advertisement

calibration mode.P1,P2 are used in low current range. P3, P4 are used in high
current range.
Command Syntax
Parameters
Examples
Related Commands

CALibrate:CURRent[:LEVel]

This command is only used in calibration mode. Input the calibration current
value according to the external measuring meter. But you must select a
calibration grade as a reference to the value you just input. These constants do
not exist in the nonvolatile memory before you send the save command
CALibrate:SAVE.
Command Syntax
Parameters
Unit
Examples
Related Commands

CALibrate:CURRent:METEr:POINt

This command is used to set the calibration point in CC mode, only effective in
calibration mode.P1,P2 are used in low current range. P3, P4 are used in high
current range. This can only calibrate the ammeter that selected through
command CONF:CURR.
Command Syntax
Parameters
Examples
Related Commands

CALibrate:CURRent:METEr[:LEVel]

This command is only effective in calibration mode. Input the calibration current
value according to reading of external measuring meter. Before you input this
value, please select a calibration grade firstly (using the command
CAL:CURR:METE:POIN).These constant do not exist in the nonvolatile
memory unless you send the command CALibrate:SAVE.
Command syntax
Parameters
Unit
Examples
Related Commands

CALibrate:VOLTage:POINt

This command is only effective in calibration mode, is used to set the calibration
point in CV mode.P1, P2 are used in low voltage range,P3,P4 are used in high
voltage range. Only be used to calibrate voltage source and voltmeter.
Command Syntax
Parameters
Examples
Related Commands
CALibrate:CURRent:POINt <point>
P1 | P2 | P3 | P4
CAL:CURR:POIN P2
CAL:STAT
CALibrate:CURRent[:LEVel] <NRf>
<external reading>
A (amps)
CAL:CURR 3.2223
CAL:STAT
CALibrate:CURRent:METEr:POINt <point>
P1 | P2 | P3 | P4
CAL:CURR:METEr:POIN P2
CAL:STAT
CALibrate:CURRent:METEr[:LEVel] <NRf>
<external reading>
A (amps)
CAL:CURR 3.2223
CAL:STAT
CALibrate:VOLTage:POINt <point>
P1 | P2 | P3 | P4
CAL:VOLT:POIN P2
CAL:STAT
Copyright © ITECH Electronic Co., Ltd.
Calibration Commands
CAL:SAV
CAL:SAV
CAL:SAV
CAL:SAV
CAL:SAV
62