Chapter12 Calibration Commands

CALibration Subsystem
Calibration commands' function:

CALibrate:SECure[:STATe]

This command can enable or disable the calibration mode. You must enable
calibration mode so that you can receive other calibration commands. The first
parameter defines enable or disable status. The second parameter represents
the code. If the calibration mode is enabled while the current code is not 0,in
this circumstance, you need to configure the second parameter. If the code is
not set or incorrect, a error will be generated and calibration mode remains
disabled. Query command can only return the status of calibration mode but
code. Everytime, when calibration status changed from enabled status to
disabled status, any new calibration constants will be lost after power-down
unless you have utilized the save command CALibrate:SAVE.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters <NR1>
Related Commands

CALibrate:INITial

This command only be used in calibration mode. It re-saves the factory
calibration constants in nonvolatile memory.
Command Syntax
Parameters
Examples
Related Commands

CALibrate:SAVe

This command only be used in calibration mode. Save the new calibration
constants in nonvolatile memory(after finishing the voltage or current calibration
process).
Command Syntax CALibrate:SAVE
Parameters
Examples
Related Commands

CALibrate:CURRent:POINt

This command is used to set the calibration point in CC mode, only effective in
Enable or disable calibration mode
Calibrate input function, current offset or gain, and save the newest
calibration constant in nonvolatile memory.
0 | 1 | OFF | ON [,<password>]
ON
CAL:SEC 0, N3301A CAL:SEC ON
CALibrate:SECure[:STATe]?
None
CAL:INIT
None
CAL:SAVE
Copyright © ITECH Electronic Co., Ltd.
CALibrate:SECure[:STATe] <bool> [,<SRD>]
CAL:SAVE CAL:INIT
CALibrate:INITial
CAL:STAT
CAL:INIT
CAL:STAT
CAL:INIT
Calibration Commands
61

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents