Dc Current Measure Verification; Dc Voltage Measure Verification; Dc Voltage Measure Verification (25.0000 To 0.0000 V) - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

DC Current Measure Verification

1. Push  key on the 715 Calibrator. Display should indicate closely to:
2. Push the key and the display should change to:
3. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
4. Connect a test lead from Black AUX terminal of the 5520A to the 715 Calibrator
5. Connect a test lead from I- terminal of the Fluke 8508A to the 715 Calibrator mA
6. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
7. Verify that the display readings on the 715 Calibrator are within the limits.

DC Voltage Measure Verification

1. Push the  key on the 715 Calibrator. Display should change to:
2. Connect test leads from the output NORMAL jacks of the Fluke 5520A to the voltage
3. Set the Fluke 5520A for the voltage settings in Table 60, and verify the display
4. Push the  key on the 715 Calibrator. Display should indicate closely to:
5. Set the Fluke 5520A for the mV settings in Table 61, and verify display readings on
sales@GlobalTestSupply.com
INPUT
-25.00 mA %
INPUT
0.000 mA
terminal.
Com terminal.
terminal.
same as the 5520A values shown in Table 59.
Table 59. 715 DC Current Measure Verification
Fluke 5520A
24.0000 mA
12.0000 mA
4.0000 mA
INPUT
0.000 V
jacks on the 715 Calibrator (black to COM jack and red to the V jack).
readings on the 715 Calibrator.
Table 60. 715 DC Voltage Measure Verification (25.0000 to 0.0000 V)
Fluke 5520A
20.00000 V
10.00000 V
0.0000 V
INPUT
0.00 mV
the 715 Calibrator.
Fluke-Direct
Fluke 715
23.993 mA to 24.007 mA
11.996 mA to 12.004 mA
3.998 mA to 4.002 mA
Fluke 715
19.996 V to 20.004 V
9.997 V to 10.003 V
-0.002 V to +0.002 V
.com
Process Calibrators
Verification
39