712 Calibration (V2.0 And Later) - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

71X Series
Calibration Manual
18. Push . The display should read:
19. Put the Fluke 8508A in the 10 kΩ range.
20. Push  and  until the 712 display matches the Fluke 8508A.
21. Push . The display should read:
22. Push  and  until the 712 display matches the Fluke 8508A.
23. Push . The display should read:
24. Push the power key to turn the unit off and disconnect the test leads.

712 Calibration (V2.0 and Later)

1. Hold down  and  while turning the power on.
2. The display shows CAL for 1 second and then xxx, where xxx is the number the
3. The calibration will be performed in the following order:
62
sales@GlobalTestSupply.com
OUTPUT ----- Ω
then change to:
OUTPUT 126.00 Ω
OUTPUT ----- Ω
then change to:
OUTPUT 2535.0 Ω
OUTPUT ----- Ω
then the unit will reset.
times the unit has been through a calibration adjustment. When the two keys are
released, PASS is displayed. One second later a zero is displayed for the most
significant of the three digit password. Use the two small arrows to change the
number. After setting each digit, push
will go into calibration. If the password was wrong, the 712 will go into normal
operation mode. The password is 217.
OHMS INPUT
OHMS OUTPUT low range
OHMS OUTPUT high range
Fluke-Direct
. If the password entry was correct, the 712
.com
1.888.610.7664