Language Settings

Users can select the language supported by this unit. The Product now support English and Chinese
UI language.
To change UI language:
1. In the Instrument Setup screen, use the up and down arrow (M N) buttons on the front panel to
move the cursor to the Language option (the option is highlighted in yellow).
2. Press D (Edit) on the front panel and select the corresponding option using the up/down
arrow buttons (M N).
3. Press D (OK) to confirm the selection and return to Instrument Setup screen.
If you press E (Cancel), the Instrument returns to the Instrument Setup screen and does not
save any change.

Max Measure Current

To measure ACR, the Instrument will output the measurement current. This option is used to set the
Maximum measurement current for 3 mΩ measurement range. A higher measurement current can
improve the signal-to-noise ratio in the measurement and obtain a higher accuracy and stable ACR
measurement value. Please see the section
To setup the maximum measurement current:
1. In the Instrument Setup screen, use the up and down arrow (M N) buttons on the front panel to
move the cursor to the Max Measure Current option (the option is highlighted in yellow). Press
D (Edit) button on the front panel to enter the maximum current setting screen. As shown in
Figure
24.
The Product provides three options for the maximum measurement current: 100 mA, 200 mA,
and 300 mA.
2. Use the up/down arrow (M N) buttons to select the appropriate option and press D (OK)
softkey to return to the Instrument Setup screen.
50
Measurement
Figure 24. Setting the max measurement current
Function.