Firmware Earlier Than V2.0; 715 Specifications - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

715 Specifications

715 Calibrator specifications vary based on the version of the instrument. To display the
firmware version for your instrument, start with the 715 off, push and hold , then push
the power button. Find the section heading below for the displayed version and use the
specification tables in that section to test and calibrate the instrument.

Firmware Earlier than V2.0

Input impedance: 2 MΩ (nominal), < 100 pF
Overvoltage protection: 30 V
Voltage drive capability: 1 mA
Overload protection: 125 mA, 250 V fast acting fuse
mA Output: 0 % = 4 mA, 100 % = 20 mA
Temperature Coefficient
0.005 % of ohms range per °C for temperature ranges – 10 °C to 18 °C and 28 °C to
55 °C
Source mode
Compliance: 1000 Ω at 20 mA for battery voltage ≥ 6.8 V (700 Ω at 20 mA for battery
voltage 5.8 to 6.8 V)
Simulate mode
External loop voltage requirement: 24 V nominal, 30 V maximum, 12 V minimum
Loop Power
24 V ±10 %
sales@GlobalTestSupply.com
Table 15. 715 DC V Input and Output
Range
100 mV
10 V
Table 16. 715 DC mA Input and Output
Range
24 mA
Fluke-Direct
Resolution
±(% of Reading + Counts)
0.01 mV
0.001 V
Resolution
±(% of Reading + Counts)
0.001 mA
Process Calibrators
Specifications
Accuracy,
0.02 % + 2
0.02 % + 2
Accuracy,
0.02 % + 2
.com
1.888.610.7664
13

Advertisement

Table of Contents
loading

Table of Contents