DC Current Amplitude Verification

1. Connect the current outputs of the Product to the current inputs of the
reference multimeter and later to the appropriate A40B current shunt.
2. Perform the dc current amplitude verification for all currents that can be
measured directly by the 200 mA range of the multimeter.
3. Perform the dc and ac current amplitude verification with the A40B current
shunt that gives the largest voltage output within the limits of the current
shunt. The A40B is used in conjunction with the reference multimeter in the
voltage mode.
Function
IDC
5-6
Range
300 mA
300 mA
300 mA
300 mA
300 mA
300 mA
300 mA
300 mA
300 mA
300 mA
300 mA
1 A
1 A
1 A
1 A
2 A
2 A
2 A
2 A
5 A
5 A
5 A
5 A
10 A
Value
30 mA
80 mA
110 mA
160 mA
210 mA
260 mA
300 mA
-30 mA
-300 mA
10 mA
-10 mA
0.31 A
-0.31 A
1 A
-1 A
1.1 A
1.8 A
-1.1 A
-1.8 A
2.1 A
5 A
-2.1 A
-5 A
5.1 A
Deviation Allowed
35.25 µA
44.0 µA
49.3 µA
58.0 µA
66.8 µA
75.5 µA
82.5 µA
35.25 µA
82.5 µA
31.75 µA
31.75 µA
0.1543 mA
0.1543 mA
0.275 mA
0.275 mA
0.393 mA
0.515 mA
0.393 mA
0.515 mA
0.868 mA
1.375 mA
0.868 mA
1.375 mA
2.571 mA