Current Verification; Dc Current Verification; Dc Voltage Measurement Verification - Fluke 5820A Service Manual

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

2. Using the N to BNC cable, connect the CHAN 1 connector on the Calibrator
Mainframe to the Fluke 5520A Calibrator voltage binding posts using a BNC to
banana connector.
Record the 5820A reading for each voltage in Table 3-10.
Nominal Value
(V dc)
0 V
1 V
2.5 V
5 V
6.5 V
9.9 V
-1 V
-2.5 V
-5 V
-6.5 V
-9.9 V

3-33. Current Verification

This procedure uses the following equipment:
Hewlett-Packard 3458A Digital Multimeter with cable
1
10
Remove the front panel Current Loop using the 2 mm hex head driver.

3-34. DC Current Verification

1. Connect the DMM directly to the vacant banana jacks on the calibrator.
2. Program the Calibrator Mainframe to output the Currents listed in Table 3-11. Press
 on the Calibrator Mainframe to activate the output.
3. Allow the HP 3458A reading to stabilize, then record the HP 3458A reading for each
current in Table 3-11.
4. Compare result to the tolerance column.
Table 3-10. DC Voltage Measurement Verification
Measured Value
100
and 1000
precision resistor assemblies
Calibration and Verification
1-year Spec.
Deviation
Verification
(V p-p)
0.001
0.0015
0.00225
0.0035
0.02625
0.03475
0.0015
0.00225
0.0035
0.02625
0.03475
3-27
3