712 Calibration (V1.1 And Earlier); Millivolts Measure - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

712 Calibration (V1.1 and Earlier)

Millivolts Measure

1. Hold down the
2. When the display momentarily shows CAL , release both the
3. Connect the test lead from the NORMAL jacks of the Fluke 5520A to the INPUT
4. Set the Fluke 5520A to 0.0000 mV and push
5. Push any key on the 712 Calibrator. The display should read:
6. Set the Fluke 5520A to 160.0000 mV.
7. Push any key on the 712 Calibrator. The display should read:
8. Set the Fluke 5520A to 320.0000 mV.
9. Push any key on the 712 Calibrator. The display should read:
10. Set the Fluke 5520A to 640.000 mV.
11. Push any key on the 712 Calibrator. The display should read:
12. Set the Fluke 5520A to 1280.000 mV.
13. Push any key on the 712 Calibrator. The display should read:
sales@GlobalTestSupply.com
and the
key.
This puts the 712 into Cal Mode. The 712 should display:
INPUT
0.00 mV
jacks (two middle jacks) on the 712 Calibrator (black to black, red to red).
Calibrator. The display should read:
INPUT
----- mV
then change to:
INPUT
0.00 mV
INPUT ----- mV
then change to:
INPUT 160.00 mV
INPUT ----- mV
then change to:
INPUT 320.00 mV
INPUT
----- mV
then change to:
INPUT
640.00 mV
INPUT ----- mV
then change to:
INPUT 1280.00 mV
INPUT ----- mV
then change to:
INPUT 400.00 R 4W
Fluke-Direct
key at the same time, then push and release the
. Push any key on the 712
Process Calibrators
and the
keys.
.com
1.888.610.7664
Calibration
59