714 Calibration (V2.0 And Later); Mv Output; Mv Input - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

6. Push any key on the 714 Calibrator. The 714 should reset itself, displaying all
7. Push the  key to turn the calibrator off and remove the Type J thermocouple test

714 Calibration (V2.0 and Later)

1. On the 714 Calibrator push and hold the
2. The display shows CAL for 1 second and then xxx, where xxx is the number the
3. The calibration will be performed in the following order:

mV OUTPUT

1. In mV OUTPUT mode, the unit will display "C -10" for 1 second followed by " –
2. Adjust the source value using the up and down big arrows for gross and small arrows
3. Push
4. Adjust the source value until you read 75.000 mV on the DMM.
5. Push

mV INPUT

1. In mV INPUT mode, the unit will display "C -10" for 1 second and then will display
2. Source -10.000 mV from the 5520. The unit should read approximately –10.000 mV
3. Push
4. Source 75.000 mV from the 5520. The unit should read approximately 75.000mV .
sales@GlobalTestSupply.com
TYPE J ------ °C
then change to:
0.xx °C
TYPE J
This number represents the absolute error of the "CJC" sensor and should be below
1°C.
segments momentarily. The display should then change to:
xx.x °C
INPUT
where xx.x is some variable number; or OL (overload) may be indicated.
lead.
pressing and releasing the  key.
times the unit has been through a calibration adjustment. When the two keys are
released, PASS is displayed. One second later a zero is displayed for the most
significant of the three digit password. Use the two small arrows to change the
number. After setting each digit, push the °C/°F key. If the password entry was
correct, the 714 will go into calibration. If the password was wrong, the 714 will go
into normal operation mode. The password is 417.
mV OUTPUT
mV INPUT
CJC
10.000". The calibrator is now sourcing mV using the default calibration. The 714
should source approximately –10.000mV through copper wire.
for fine, until you read -10.000mV.
key to go to the next calibration point. "C 75" will be displayed for 1
second and then " 75.000". The 714 should source approximately 75.000mV.
to complete mV OUTPUT calibration.
the input voltage using the default calibration.
through copper wire.

to go to the next calibration point. "C 75" will be displayed for 1 second
and then the input voltage.
Fluke-Direct
and
keys, then turn on the 714 by
.com
Process Calibrators
Calibration
65