718 and 718Ex Specifications

Accuracy is specified for 1 year after calibration at operating temperatures of -10 °C to +
55 °C.
To display firmware version, start with the unit off, push and hold  or  then push
.
Model
718-1G
718-30G
718-100G
718-300G
Accuracy: Pressure Accuracy is ±0.05% of positive range
Temperature coefficient: .01 % of range per °C for temperature ranges of between −10 °C to 18 °C and
28 °C to 55 °C.
Model
718-1G
718-30G
718-100G
718-300G
[1] In an RF field of 1 V/m to 3 V/m between the frequencies of 1.4 GHz to 2.0 GHz, add 120 counts to
specified accuracy.
18
sales@GlobalTestSupply.com
Table 23. Pressure Specifications
718 Ex Versions Under 1.9
Range SI
(-1 to 1) PSI
(-7 to 7) kPa
(-12 to 30) PSI
(-83 to 207) kPa
(-12 to 100) PSI
(-83 to 690) kPa
(-12 to 300) PSI
(-83 to 2068) kPa or 2.1 mPa
Range
-1 to 1 PSI
5 PSI
-7 to 7 kPa
34.5 kPa
-12 to 30 PSI
60 PSI
-83 to 207 kPa
413 kPa
-12 to 100 PSI
200 PSI
-83 to 690 kPa
1.4 mPa
-12 to 300 PSI
375 PSI
-83 to 2.1 mPa
2.6 mPa
Table 24. Pressure Display, Pressure Module Input
Range
Refer to the Instruction Sheet for the pressure module
Fluke-Direct
718 Version 1.3 to 3.9
Range Metric
718 Version 4.0 or Higher
718 Ex V 2.0 and Higher
Max
Resolution
Max SI
Max Metric
5 PSI
34.5 kPa
60 PSI
413 kPa
200 PSI
1379 kPa or 1.4 mPa
375 PSI
2586 kPa or 2.6 mPa
Accuracy [1]
Pressure accuracy is ± % of
positive range
6 month
1 year
0.050 %
0.050 %
0.025 %
0.035 %
0.025 %
0.035 %
0.035 %
0.050 %
Accuracy