Resistance Measure; Ma Measure - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

71X Series
Calibration Manual

Resistance Measure

1. Connect two additional test leads from the NORMAL jacks of the 5520A to the 712
2. Set the Fluke 5520A to 400.00 Ω and push
3. Push any key on the 712 Calibrator. The display should read:

mA Measure

1. Disconnect all four test leads from the Fluke 5520A and the 712 Calibrator.
2. Connect the two test leads from the AUX jacks on the Fluke 5520A to the middle
3. Push any key on the 712 Calibrator. The display will change to:
4. Set the Fluke 5520A to 0.00000 mA and push
5. Push any key on the 712 Calibrator. The display should read:
6. Again, push any key on the 712 Calibrator, still with a 0.00000 mA input from the
7. Set the Fluke 5520A to 2.00000 mA.
8. Push any key on the 712 Calibrator. The display should read:
9. Set the Fluke 5520A to 0.50000 mA.
10. Push any key on the 712 Calibrator. The display should read:
11. Push any key on the 712 Calibrator and the unit will automatically reset itself, with
60
sales@GlobalTestSupply.com
Calibrator (black wire to the 712 red jack labeled [3W NC] and red wire to the red
jack labeled [4W NC]. All four input jacks of the 712 should now be in a 4-wire
ohms measurement configuration.
INPUT ----- R 4W
then change to:
INPUT -----
jacks on the 712 Calibrator (black to black; red to red).
INPUT
0.00 mA
INPUT ----- mA
then change to:
INPUT 0.00 mA
Fluke 5520A. The display should read:
INPUT ----- mA
then change to:
INPUT 2.000 mA
INPUT ----- mA
then change to:
INPUT 0.500 mA
INPUT ----- mA
all segments momentarily being displayed. Calibrator is now out of CAL mode. Push
key to turn the calibrator off and disconnect all test leads.