714 English Instruction Sheet
Thermocouple Calibrator
Introduction
The Fluke 714 Thermocouple Calibrator is a precise source and
measurement tool for calibrating thermocouple instruments. The
calibrator sources or measures in units of °C, °F, or mV, through
a thermocouple minijack.
Your calibrator is supplied with a Flex-Stand™ holster, an
installed 9 V alkaline battery, and this instruction sheet. Sets of
thermocouple miniplugs are available from Fluke. (Accessories
Fluke-700TC1 and Fluke-700TC2 TC Miniplug Kits.)
If the calibrator is damaged or something is missing, contact the
place of purchase immediately. Contact your Fluke distributor for
information about accessories. To order replacement parts or
spares, see "Replacement Parts."
The following tables list the thermocouple types supported by the
calibrator, the standards and scales used for each type, the
thermocouple properties, and calibrator resolution. Full calibrator
specifications are listed at the end of this instruction sheet.
Since mV input and output units are available, you can
use the calibrator for any thermocouple type by making
manual calculations or referring to tables.
PN 650306 July 1997 Rev. 1, 10/97
©1997 Fluke Corporation. All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.
Fluke-Direct
Instruction Sheet
Note
Page 1
®
714
.ca
1.800.868.7495

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 714

  • Page 1 ® Thermocouple Calibrator Instruction Sheet Introduction The Fluke 714 Thermocouple Calibrator is a precise source and measurement tool for calibrating thermocouple instruments. The calibrator sources or measures in units of °C, °F, or mV, through a thermocouple minijack. Your calibrator is supplied with a Flex-Stand™ holster, an installed 9 V alkaline battery, and this instruction sheet.
  • Page 2 714 English Instruction Sheet Page 2 Thermocouple Standards and Scales Thermocouple Type Standard Scale J, K, T, E, R, S, B NIST 175 ITS-90 L (J-DIN), U (T-DIN) DIN 43710 IPTS-68 Thermocouple Properties Thermocouple Temperature Display Resolution Type Ranges 0.1°C or °F -200 to 1200°C,...
  • Page 3: Explanation Of International Symbols

    714 English Instruction Sheet Page 3 Millivolt Range and Resolution Mode Range Display Resolution -10 to 75 mV 0.01 mV Explanation of International Symbols The following symbols are used on the calibrator or in this instruction sheet. The table below explains their meaning.
  • Page 4: Safety Information

    714 English Instruction Sheet Page 4 Safety Information W Warning To avoid possible electric shock or personal injury: • Never apply more than 30 V between the TC terminals, or between either TC terminal and earth ground. • Make sure the battery door is closed and latched before you operate the calibrator.
  • Page 5: Measuring A Thermocouple

    714 English Instruction Sheet Page 5 Measuring a Thermocouple INPUT Shows TC type selected THERMOCOUPLE CALIBRATOR Press to select Press to toggle °C or °F TC type or mV Press so that INPUT is on the display TC wire and...
  • Page 6: Simulating A Thermocouple

    714 English Instruction Sheet Page 6 Simulating a Thermocouple OUTPUT Shows TC type selected THERMOCOUPLE CALIBRATOR Press to select Press to toggle °C or °F TC type or mV Press so that OUTPUT is on the display Press to scroll up/down 0.1°...
  • Page 7: Maintenance

    If the calibrator needs repair, contact a Fluke Service Center. If the calibrator is under warranty, see the warranty statement for terms. If the warranty has lapsed, the calibrator will be repaired and returned for a fixed fee. Contact a Fluke Service Center for information and price. Cleaning Periodically wipe the case with a damp cloth and detergent;...
  • Page 8: Replacing The Fuse

    714 English Instruction Sheet Page 8 it07f.eps Replacing the Fuse ! Warning To avoid personal injury or damage to the calibrator, use only a 0.125A 250V fast fuse, ® Littelfuse 2AG. Fuse F1 is probably blown if in the input mode, the calibrator always reads OL, even with a thermocouple connected.
  • Page 9 714 English Instruction Sheet Page 9 Replace the fuse as follows: Remove the test leads and turn the calibrator off. Remove the battery door. Remove the three Phillips-head screws from the case bottom and turn the case over. Gently lift the top cover from the end nearest the input/output terminals until it unsnaps from the bottom cover.
  • Page 10 714 English Instruction Sheet Page 10 Snaps ke03f.eps Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 11: Replacement Parts And Accessories

    714 English Instruction Sheet Page 11 Replacement Parts and Accessories Replacement Parts Item Description PN or Qty. Model 9V battery, ANSI/NEDA 1604A or 614487 IEC 6LR61 CG81Y Holster, Yellow CG81Y Fuse, 125 mA, 250V fast 686527 MP85 Case top 620234...
  • Page 12 714 English Instruction Sheet Page 12 MP85 MP86 H2, 3, 4 MP89, 90 H5, 6 MP92 Instruction Sheet Holster ke04c.eps Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 13: Specifications

    714 English Instruction Sheet Page 13 Specifications Specifications are based on a one year calibration cycle and apply for ambient temperature from +18°C to +28°C unless stated otherwise. “Counts” means number of increments or decrements of the least significant digit.
  • Page 14: General Specifications

    714 English Instruction Sheet Page 14 General Specifications Maximum voltage applied between any terminal and earth ground or between any two terminals: 30 V Storage temperature: -40°C to 60°C Operating temperature: -10°C to 55°C Operating altitude: 3000 meters maximum Temperature coefficient: 0.05 x specified accuracy per °C for temperature ranges -10°C to 18°C and 28°C to 55°C...
  • Page 15 Page 15 LIMITED WARRANTY & LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling.