Specifications - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

71X Series
Calibration Manual

Specifications

Specifications for the 71X Calibrators are based on a one-year calibration cycle and apply
for ambient temperatures from + 18 °C to + 28 °C unless stated otherwise. "Counts" are
the number of increments or decrements of the least significant digit. General
specifications for all models are in Table 2.
Maximum voltage applied between any terminal
and earth ground or between any two terminals:
Storage temperature:
(718Ex) Pressure Sensor Media:
Operating temperature:
Operating altitude:
Relative humidity:
Vibration:
Shock:
Safety: 712, 714, 715, 717
Safety: 718
Safety: 718Ex
719
Power requirements: 712, 714, 715, 717
Power requirements: 718
Power requirements: 718Ex
Size: 712, 714, 715, 717
Size: 718, 718Ex, 719
Weight:
712
714
715
717 30G, 717 100G
718 30G and 718 100G and
718Ex 30G and 718Ex 100G
719
6
sales@GlobalTestSupply.com
Table 2. General Specifications
30 V
712: -20 °C to 60 °C, 718Ex: -40 °C to 71 °C, 719: -30 °C to 60 °C
All other models: -40 °C to 60 °C
Non-corrosive gasses only
-10 °C to 55 °C
3000 meters maximum
95 % up to 30 °C
75 % up to 40 °C
45 % up to 50 °C
35 % up to 55 °C
Random 2 g, 5 Hz to 500 Hz
1 meter drop test
Certified as compliant to CAN/CSA C22.2 No. 1010.1:1992 NRTL
Complies with ANSI/ISA S82.01-1994
Certified as compliant to CAN/CSA C22.2 No. 1010.2:1995
Complies with ANSI/ISA S82.01-1995
Certified as compliant to CAN/CSA C22.2 No. 1010.2:1995
Complies with ANSI/ISA S82.01-1995. Complies with IEC 61010-1-
95 CAT I, 30 V;
P
0344
)
LR110460
Ta = -10 °C... +55 °C
CE: Complies with EN61010-1 and EN61326
Complies with EN/IEC 61010-1 2
CSA-C22.0 No. 61010-1-04
Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
Two 9 V batteries (ANSI/NEDA 1604A or IEC 6LR61)
See "718Ex Approved Batteries".
34.9 mm H x 87 mm W x 187 mm L;
With holster and Flex-Stand:
52 mm H x 98 mm W x 201 mm L
60 mm H x 87 mm W x 210 mm L;
With holster: 66 mm H x 94 mm W x 216 mm L
337 g; With holster and Flex-Stand: 587 g
332 g; With holster and Flex-Stand: 584 g
349 g; With holster and Flex-Stand: 601 g
369 g; With holster and Flex-Stand: 624 g
737 g; With holster: 992G
912 g; With holster
Fluke-Direct
(
II 1 G Ex ia IIC T4
KEMA 04ATEX1061 X
Class I Div. 1 Groups A-D T4
AEx ia IIC T4
nd
ed.,
.com
1.888.610.7664

Advertisement

Table of Contents
loading

Table of Contents