Ohms Input; Ohms Output Low Range; Ohms Output High Range - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

OHMS INPUT

1. In OHMS INPUT mode, the unit will display "C 15" for 1 second and then display
2. Connect the test leads from the 5520 main Jacks to the 712 jacks in 4 wire
3. Push
4. Source 400 Ω from the 5520. The unit should read 400.0 Ω ±5%.
5. Push
6. Source 3200 Ω from the 5520. The unit should read 3200.0Ω ±5%.
7. Push
8. Push

OHMS OUTPUT Low Range

1. In OHMS OUTPUT low range mode, the unit will display " 15.00" and will source
2. Connect the 712 to the DMM in OHMS read using 4 wire mode connection (2 wires
3. Adjust the value for OHMS OUTPUT using the arrows on the 712 keypad to source
4. Push
5. Adjust using the arrows on the keypad to source exactly 400.00 Ω.
6. Push
7. Push

OHMS OUTPUT High Range

1. In OHMS OUTPUT high range mode, the unit will display " 20.0" and will source
2. Adjust the value for OHMS OUTPUT using the arrows on the 712 keypad to source
3. Push
4. Adjust using the arrows on the keypad to source exactly 4000.0 Ω.
5. Push
6. Push
sales@GlobalTestSupply.com
the OHMS input using the default calibration.
configuration (2 wires at 5520, 4 wires at 712). The unit should read 15.0 Ω ±5%.

to go to the next calibration point. "C 400" will be displayed for 1 second
and then it will display the ohms input.

to go to the next calibration point. "C3200" will be displayed for 1 second
and then it will display the ohms input.

to complete OHMS INPUT calibration. "CHEC" will be displayed for 1
second. The new calibration constants are applied and the OHMS INPUT calibration
can be checked.

to go to the next mode.
OHMS using the default calibration.
at 712, 4 wires at DMM). The DMM should read 15.00 Ω ±5%.
exactly 15.00 Ω.
to go to the next calibration point. "400.00" will be displayed

to complete OHMS OUTPUT low range calibration. "CHEC" will be
displayed for 1 second. The new calibration constants are applied and the calibration
can be checked.

to go to the next mode.
OHMS using the default calibration. Keep the connections from the previous mode.
The DMM should read 20.0 Ω ±5%.
exactly 20.0 Ω.

to go to the next calibration point. "4000.0" will be displayed.
to complete OHMS OUTPUT high range calibration. "CHEC" will be
displayed for 1 second. The new calibration constants are applied and the calibration
can be checked.
to complete calibration.