DC Voltage Measure Verification

1. Push the  key on the 715 Calibrator. Display should change to:
2. Connect test leads from the output NORMAL jacks of the Fluke 5520A to the voltage
3. Set the Fluke 5520A for the voltage settings in Table 54, and verify the display
4. Push the  key on the 715 Calibrator. Display should indicate closely to:
5. Set the Fluke 5520A for the mV settings in Table 55, and verify display readings on
6. Disconnect the test leads from 715 Calibrator and turn the green  key off.

Display Verification

1. Push and hold the  key and then turn the 715 back on by pressing and releasing
2. Check to see that all segments of the LCD are displayed.
3. Turn the 715 off.
sales@GlobalTestSupply.com
INPUT
0.000 V
jacks on the 715 Calibrator (black to COM jack and red to the V jack).
readings on the 715 Calibrator.
Table 54. 715 DC Voltage Measure Verification (10.0000 to 0.0000 V)
Fluke 5520A
10.00000 V
5.00000 V
0.0000 V
INPUT
0.00 mV
the 715 Calibrator.
Table 55. 715 DC Voltage Measure Verification (0.0000 mV to 100.0000 mV)
Fluke 5520A
0.0000 mV
50.0000 mV
100.0000 mV
the green  key. This locks the 715 in a mode where all display segments are on. All
segments will stay on until the  key is released.
The 715 verification test is complete.
Fluke-Direct
Fluke 715
9.996 V to 10.004 V
4.997 V to 5.003 V
-0.002 V to +0.002 V
Fluke 715
-0.02 mV to + 0.02 mV
49.97 mV to 50.03 mV
99.96 mV to 100.04 mV
.com
Process Calibrators
Verification
37