717 Calibration (Earlier than V2.0)

mA Measure

1. Hold down the  and the  keys at the same time, then push the green
2. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
3. Set the Fluke 5520A to 0.00000 mA and push
4. Push any key on the 717 Calibrator. The display should change to:
5. Set the Fluke 5520A to 24.0000 mA.
6. Push any key on the 717 Calibrator. Display should change from:

Pressure Measure

1. Remove test leads from the 717 Calibrator and carefully attach the pressure fitting to
The use of TEFLON tape at the pressure fittings strengthens the seal.
2. For the 717 100G push the  key. For the 717 30G, push the
3. Vent the system to ambient air to ensure 0.00 PSI.
4. Push any key on the 717 Calibrator. The display will change to read:
5. Set up the deadweight tester for 30 PSI (or 100 PSI) to be injected into the pressure
70
sales@GlobalTestSupply.com
When the display shows CAL momentarily, release the  and the  keys
immediately. After CAL shows, the display will automatically read:
CAL
0.000mA
the 717 Calibrator (Black to Black and Red to Red).
-------
0.000mA
then to:
CAL
24.000mA
-------
24.000mA
to:
rAnGE
the pressure jack on the 717 Calibrator.
will flash for a short time:
------
rAnGE
then display:
CAL
0.00 PSI
CAL
------ PSI
then depending on the model selected in step 2, change to either:
CAL
30.000 PSI
or
CAL
100.00 PSI
port of the 717 Calibrator.