Ma Measure Verification - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

71X Series
Calibration Manual
Input Pressure
0 PSI
2000 PSI
4000 PSI
6000 PSI
8000 PSI
10000 PSI
8000 PSI
6000 PSI
4000 PSI
2000 PSI
*To display the firmware version, start with the 717 off, push and hold the center bottom button ( or ),
then push the power button.
4. Carefully vent all pressure and disconnect the 717 from the deadweight tester.

mA Measure Verification

1. Push
2. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
3. Connect a test lead from Black AUX terminal of the 5520A to the 717 Calibrator
4. Connect a test lead from I- terminal of the Fluke 8508A to the 717 Calibrator mA
5. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
6. Verify that the display readings on the 717 Calibrator are within the limits.
Fluke 5520A
4.0000 mA
12.0000 mA
24.0000 mA
48
sales@GlobalTestSupply.com
Table 62. 717 Pressure Verification (cont.)
6-Month Display Reading
-4 to 4
1996 to 2004
3996 to 4004
5996 to 6004
7996 to 8004
9996 to 10004
7996 to 8004
5996 to 6004
3996 to 4004
1996 to 2004
to turn power on. The display should read:
----------
mA
then change to:
0.000 mA
0.00PSI
terminal.
Com terminal.
terminal.
same as the 5520A values shown in Table 63.
Table 63. 717 mA Measure Verification
Fluke 717 (V2.9 and lower)
3.998 mA to 4.002 mA
11.996 mA to 12.004 mA
23.993 mA to 24.007 mA
Fluke-Direct
Version 4.0 or Higher
717 10000G
-5 to 5
1995 to 2005
3995 to 4005
5995 to 6005
7995 to 8005
9995 to 10005
7995 to 8005
5995 to 6005
3995 to 4005
1995 to 2005
1-Year Display Reading
Fluke 717 (V3.0 and higher)
3.997 mA to 4.003 mA
11.996 mA to 12.004 mA
23.994 mA to 24.006 mA