Performance Tests; Verification Tests For Dc Voltage (Normal) - Fluke 5080A Operator's Manual

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

Performance Tests

To verify that the 5080A meets its specifications, you can use Tables 7-3 through 7-12.
The tables are for qualified metrology personnel who have access to a standards
laboratory that is properly equipped to test calibration equipment of this level of
accuracy. The tables show the recommended test points and the acceptable upper and
lower limits for each point. The limits were computed simply by adding or subtracting
the 90-day specification from the output value. There is no built-in factor for
measurement uncertainty. If you need more detailed verification and calibration
adjustment instructions, order the 5080A Service Manual.
Table 7-3. Verification Tests for DC Voltage (Normal)
Range
330 mV
0V .3V range
330 mV
0.3V .3V range
330 mV
-0.3V .3V range
3.3 V
0V 3V range
3.3 V
0.3V 3V range
3.3 V
-0.3V 3V range
3.3 V
1V 3V range
3.3 V
-1V 3V range
3.3 V
3V 3V range
3.3 V
-3V 3V range
33 V
0V 30V range
33 V
30V 30V range
33 V
-30V 30V range
330 V
30V 300V range
330 V
-30V 300V range
330 V
300V 300V range
330 V
-300V 300V range
1000 V
100V 1000V range
1000 V
-100V 1000V range
1000 V
1000V 1000V range
1000 V
-1000V 1000V range
Output
Nominal
0
0.3
-0.3
0
0.3
-0.3
1
-1
3
-3
0
30
-30
30
-30
300
-300
100
-100
1000
-1000
Lower Limit
Upper Limit
-0.00001 mV
0.00001 mV
0.299951 mV
0.300049 mV
-0.300049 mV
-0.299951 mV
-0.000015 V
0.000015 V
0.299955 V
0.300045 V
-0.300045 V
-0.299955 V
0.999885 V
1.000115 V
-1.000115 V
-0.999885 V
2.999685 V
3.000315 V
-3.000315 V
-2.999685 V
-0.00015 V
0.00015 V
29.99685 V
30.00315 V
-30.00315 V
-29.99685 V
29.9949 V
30.0051 V
-30.0051 V
-29.9949 V
299.9625 V
300.0375 V
-300.0375 V
-299.9625 V
99.9825 V
100.0175 V
-100.0175 V
-99.9825 V
999.8745 V
1000.1255 V
-1000.1255 V
-999.8745 V
7
Maintenance

Performance Tests

7-9