NDD EasyOne Air Operator's Manual page 104

Hide thumbs Also See for EasyOne Air:
Table of Contents

Advertisement

Appendix
Immunity test standard
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
 Electromagnetic immunity
Operator's Manual V03 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
IEC 60601 test level
Compliance level
3 Vrms
3 Vrms
150 kHz to 80 MHz
150 kHz to 80 MHz
3 V/m
10 V/m
80 MHz to 2.5 GHz
80 MHz to 2.7 GHz
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio
a
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
EasyOne Air is used exceeds the applicable RF compliance level above, the EasyOne Air
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
EasyOne Air.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
b
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the EasyOne Air, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d
1.2 P
=
d
0.35 P
80 MHz to 800 MHz
=
d
0.7 P
800 MHz to 2.5 GHz
=
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey,
should be less than the
compliance level in each frequency
b
range.
Interference may occur in the vicinity of
equipment marked with the following
icon:
9000007/03
105
104/105