Safety Information Regarding Electromagnetic Compatibility - NDD EasyOne Air Operator's Manual

Hide thumbs Also See for EasyOne Air:
Table of Contents

Advertisement

Safety information
Malfunction
False diagnosis
Excessive heat
Network connection
2.3.

Safety information regarding electromagnetic compatibility

Influence by HF
surgical equipment
Portable wireless
communications
equipment
Increased emissions
or decreased
immunity
Proximity to other
equipment
Operator's Manual V03 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Non-original accessories and disposables, like respiratory tubes, battery,
power adapter, or cables can cause malfunction.
Only use original accessories and disposables by the manufacturer.
List of order numbers and accessories for EasyOne Air  9 3
Non-original respiratory tubes can cause measurement error and false results.
You must only use EasyOne FlowTube respiratory tubes by the manufacturer
ndd to assure accuracy, long-life, and full warranty coverage.
Prolonged exposure to direct sunlight, for example in cars, or exposure to oth-
er sources of heat may cause the battery to explode.
Do not store in places that can exceed temperatures of 50°C or 122°F.
Remove the battery and store the battery in the original packaging.
Connection of EasyOne Connect to an IT network that includes other equip-
ment could result in previously unidentified risks to patients, operators, or
third parties.
Subsequent changes to the IT network could introduce new risks and require
additional analysis.
Changes to the IT network include:
• changes in the IT network configuration,
• connection of additional items to the IT network,
• disconnecting items from the IT network,
• update of equipment connected to the IT network, and
• upgrade of equipment connected to the IT network.
It is your responsibility to identify, analyze, evaluate, and control these risks.
IEC 80001-1:2010 provides guidance to address these risks.
Electrical devices with a high RF power output during intended use (e.g. High
frequency (HF) surgical equipment) should not be operated in parallel with
EasyOne Air.
Portable wireless communications equipment such as wireless home network
devices, mobile phones, cordless telephones and their base stations, walkie-
talkies, etc. can affect EasyOne Air.
Keep a distance of at least 30 cm (12 in) to any part of EasyOne Air.
Accessories, transducers, and cables other than those specified by the manu-
facturer or replacement parts for internal components may result in increased
emissions or decreased immunity of EasyOne Air.
EasyOne Air should not be used adjacent to or stacked with other equipment.
Should close proximity or stacking be unavoidable, however, the configura-
tion in which it is used needs to be watched closely to ensure that the equip-
ment is functioning normally.
9000007/03
16
13/105