NDD EasyOne Air Operator's Manual page 92

Hide thumbs Also See for EasyOne Air:
Table of Contents

Advertisement

Specifications and bibliography
The following standard conditions are identical to the conditions of EasyOne Air. Storage, transport, and operation under
these conditions is safe.
Standard storage and transport
conditions
Standard operating conditions
The following optimized conditions prolong the life expectancy of the battery.
Optimized environmental
conditions
Optimized operating conditions
while charging and discharging
Optimized storage periods and
storage conditions resulting in no
less than 80% recoverable capacity
of the initial capacity
Icons on the battery
 Specifications for the lithium-ion battery of EasyOne Air
Operator's Manual V03 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Temperature
-20°C to 50°C (-4°F to 122°F)
Relative humidity
5% to 90%
Atmospheric pressure
500 hPa to 1060 hPa
Battery
removed battery
Temperature
0°C to 40°C (32°F to 104°F)
Relative humidity
5% to 90%
Atmospheric pressure
700 hPa to 1060 hPa
Temperature
23±5°C (73.4±9°F)
Humidity
45% to 75% relative humidity
Atmospheric Pressure
860 hPa to 1060 hPa
Charging
0°C to 45°C (32°F to 113°F)
45% to 85% relative humidity
Discharging
-20°C to 60°C (-4°F to 140°F)
45% to 85% relative humidity
Storage period
-20°C to 60°C (-4°F to 140°F)
≤1 month
45% to 85% relative humidity
Storage period
-20°C to 45°C (-4°F to 113°F)
≤3 months
45% to 85% relative humidity
Storage period ≤1 year -20°C to 20°C (-4°F to 68°F)
45% to 85% relative humidity
The battery complies with the requirements of applicable
EC directives of the European Union.
Consult instructions for use.
Caution, consult accompanying documents.
In the European Union, the product you have purchased
should not be disposed of as unsorted municipal waste.
Please make use of your local WEEE collection facilities to
dispose of this product and otherwise observe all
applicable requirements.
The battery was certified for the US market by SGS CSTC
Standards Technical Services Co., Ltd., Shanghai in project
number 710351.
Batch code
9000007/03
94